Geniş Kapsamlı Elektriksel Analiz Test Cihazı
Product Detail
*
Integration:ACW,DCW,IR,GB,LC,PWR,STR(7 functions in 1)
* Information:barcode recognition,program
contorl,data storage,query statistics
* High quality:stable,accurate data,fast
test,easy operation
* Strong Expansion:Support ESRS software,IN-PAD
termianl and power supply
Main features
* 7 in 1:AC withstand voltage 5000V/40mA,DC
withstand voltage 6000V/10mA,insulation resistance 2500V/9999MΩ,ground
resistance 32A/600mΩ,leakage current complying with IEC60990 and Figure 4 in
GB/T12113,AN9642H also has operating power 300V/20A and low voltage start
functions.
* Fast test: if the setting time is total 5s,it
can be fastert finshed in 5.6s for all testes, The accuracy of safety test is
2%,power parameters accuracy is 0.2% and the minimum measurement power is 1w.
* Automated test: Automation and information
interface,Supporting combinging with barcode scanners,servers and other
automated testers to realize testing automatically.
* Intelligent management:Barcode
recognition,matching tes program,intelligent programmable control of the test
conditions;test data storage(local U disk), server data exchange (network
interface orRS232), data analysis (ESRS software) and other information
management;meeting the MES management requirements in modern line.
* Open extension:Rich extension,including 100mA
with stand voltage,64A ground,the measurement network selection of leakage
current and other options,optional IN-PAD,software(ESRS),barcode scanners,power
supply etc.
Spec.
AC Withstand Voltage Test | |||
Rated output |
200VA (5000V/40mA), short circuit current is greater than 200mA; [Optional: 5000V/100mA] | ||
Output voltage |
Range/accuracy |
100~5000V, ± (2%× setting + 5 words) | |
Output frequency |
Range/accuracy |
50Hz or 60Hz, ±0.1%× setting | |
Alarm current |
Upper limit/accuracy |
0.10~ or 40.00mA, ± (2%× setting + 5 words) | |
Lower limit /accuracy |
0.000~9.999mA, ± (2%× setting + 5 words) | ||
Time |
Range |
0.5~999.9s (0: infinite time) | |
Ramp time |
0.1~999.9s | ||
Descent time |
0.1~999.9s | ||
Accuracy |
± (0.2%× setting + 1 word) | ||
Measure meter |
Voltage range/accuracy |
0.01~5.00kV, ± (1.5%× reading + 1 word) | |
Frequency range/accuracy |
50Hz or 60Hz, ±0.1%× reading | ||
Current range/accuracy |
0.10~40.00mA, ± (2%× reading + 2 words) | ||
Time range/accuracy |
0.1~999.9s, ± (0.2%× reading + 1 word) | ||
Output voltage wave, distortion, adjusting rate |
Sine wave, less than 2% (pure resistive load), less than (2% × setting + 5V) (from zero to full load) | ||
Start voltage |
(0%~50%)× output voltage setting | ||
Current compensation |
0.000~40.00mA, Auto/manual | ||
Arc detection |
Level 1~9, 9: highest level; 0: off | ||
DC withstand voltage test | |||
Output voltage |
Range/accuracy |
100~6000Vdc, ± (2%× setting + 5 words) | |
Alarm current |
Upper limit/accuracy |
0~10000μA, ± (2%× setting + 5 words) | |
Lower limit /accuracy |
0.0~999.9μA, ± (2%× setting + 5 words) | ||
Time |
Range |
0.5~999.9s (0: infinite time) | |
Ramp time |
0.1~999.9s | ||
Descent time |
0.3~999.9s (0, off) | ||
Accuracy |
± (0.2%× setting + 1 word) | ||
Measure meter |
Voltage range/accuracy |
0.01~6.00Vdc, ± (1.5%× reading + 1 word) | |
Current range/accuracy |
0.0~10000uA, ± (2%× reading + 5 words) | ||
Time range/accuracy |
0.1~999.9s, ± (0.2%× reading + 1 word) | ||
Rated output |
60VA (6000Vdc/10mA) | ||
Output voltage fluctuation, regulation |
Less than 5% (purely resistive load, 6000V/10mA), less than (2% × setting + 5V) (from zero to full load) | ||
Compensation current |
0 ~200.0μA, Auto/manual | ||
Start voltage |
(0%~50%)×output voltage setting | ||
Charging current lower limit |
(0~3.500) μA, Auto/manual | ||
Discharge time |
≤200ms | ||
Arc detection |
Level 1~9, 9: highest level; 0: off | ||
Insulation resistance test | |||
Output voltage |
Range/accuracy |
100~2500Vdc, ± (2%× setting +5V) | |
Alarm resistance |
Upper/lower limit |
1~9999MΩ |